Figure 7-27: Peripheral PCA BoardPeripheral Board
A. Peripheral board
B. Ribbon cable (to motherboard)
C. Cable tie
D. TEC board ribbon cable harness
E. 12 V power wiring harness
F. M3 flat washer harness
G. M3 spring washer
H. M3 x 6 mm socket head cap screw
I. Relay wiring harness
J. Gas/cell temp monitoring
K. Gas/cell temp monitoring
L. Gas/cell temp monitoring
4. Cut the cable tie (C) that secures the TEC board ribbon cable (D) to the connector of
the peripheral board ribbon cable (B). Carefully move the TEC board ribbon cable
clear of the peripheral board.
5. Disconnect the ribbon cable (B) from the peripheral PCA board (A).
Troubleshooting and diagnostics
Reference Manual
September 2019
00809-0100-5800
126
Emerson.com/Rosemount
Summary of Contents for Rosemount CT5800
Page 20: ...Contents Reference Manual September 2019 00809 0100 5800 xx Emerson com Rosemount ...
Page 34: ...Plan Reference Manual September 2019 00809 0100 5800 34 Emerson com Rosemount ...
Page 60: ...Startup procedure Reference Manual September 2019 00809 0100 5800 60 Emerson com Rosemount ...
Page 90: ...Gas calibration Reference Manual September 2019 00809 0100 5800 90 Emerson com Rosemount ...
Page 168: ...Shutdown procedure Reference Manual September 2019 00809 0100 5800 168 Emerson com Rosemount ...
Page 178: ...Theory of operation Reference Manual September 2019 00809 0100 5800 178 Emerson com Rosemount ...
Page 180: ...Spare parts list Reference Manual September 2019 00809 0100 5800 180 Emerson com Rosemount ...