3.3
Startup procedure
NOTICE
The gases shown in the screenshots and the measurements thereof may be different from
those shown in your particular analyzer. They indicate the functionality of the software,
which is the same regardless of the gases being measured.
NOTICE
To stop the startup procedure at any time, set the main circuit breaker to
OFF
.
To start up the analyzer, perform the following steps:
Procedure
1. Visually inspect the exterior of the analyzer for signs of damage, corrosion, gas
leaks, or overheating. Report anything found to the hazardous area Ex maintenance
organization.
2. Make sure that the analyzer has been correctly installed.
See
3. Make sure that the enclosure is closed and all bolts are correctly torqued.
If the enclosure cannot be closed and locked, report it to the maintenance
organization and do not proceed further until the door has been repaired.
4. Make sure that the gas handling system is turned OFF.
5. Make sure that the external circuit breaker is set to OFF.
6. Visually examine the gas ports to make sure that the sample supply line and the
sample return line are correctly attached to the analyzer.
If necessary, refer to
Replacing the gas input and output lines
.
7. Visually check that the electrical connections have been made to the instrument.
8. Set the main circuit breaker to ON, which applies main power to the instrument.
The analyzer begins its automatic startup. The control PC that forms part of the
instrument is configured to automatically load the necessary gas sensor software
and configuration files. The startup sequence commences automatically under
software control.
After a few seconds, the
Gas Sensor Main
screen (
controller. If it does not, report the fault to the maintenance organization.
Startup procedure
Reference Manual
September 2019
00809-0100-5800
58
Emerson.com/Rosemount
Summary of Contents for Rosemount CT5800
Page 20: ...Contents Reference Manual September 2019 00809 0100 5800 xx Emerson com Rosemount ...
Page 34: ...Plan Reference Manual September 2019 00809 0100 5800 34 Emerson com Rosemount ...
Page 60: ...Startup procedure Reference Manual September 2019 00809 0100 5800 60 Emerson com Rosemount ...
Page 90: ...Gas calibration Reference Manual September 2019 00809 0100 5800 90 Emerson com Rosemount ...
Page 168: ...Shutdown procedure Reference Manual September 2019 00809 0100 5800 168 Emerson com Rosemount ...
Page 178: ...Theory of operation Reference Manual September 2019 00809 0100 5800 178 Emerson com Rosemount ...
Page 180: ...Spare parts list Reference Manual September 2019 00809 0100 5800 180 Emerson com Rosemount ...