Figure 4-5: Gas Sensor Main Screen
The Gas Sensor Main screen displays the gas concentration measurements obtained by
the analyzer. In the example shown in
, the gases carbon dioxide (CO
2
), methane
(CH
4
), carbon monoxide (CO), and water (H
2
O) are being measured; for each gas, the
concentration detected is in parts per million (ppm).
At the end of the start-up procedure, the gas measurements initially appear as
0.00 ppm
until the first readings are taken. After a few seconds, the initial gas concentrations are
displayed.
The
Gas Sensor Main
screen also shows the status of the analyzer. In the example shown
in
Running
and
OK
(i.e., no faults have been identified).
If a fault is identified,
?
is displayed; if maintenance is required,
is displayed.
To access other screens through the
Main menu
, press
).
is a link between the
Gas Sensor Main
) and the
Pressure and
Temperature
screen (
). Press
to toggle between the two screens.
is a link to the Help system. Press
to go to a
Help
On the
Gas Sensor Main
screen,
has no function when the analyzer is operating
correctly. If, however, the software detects a fault, an error message is displayed. Press
to get further information on the error.
4.6
Pressure and Temperature screen
The
Pressure and Temperature
) shows pressure and temperature
measurements taken inside the analyzer.
Operating the analyzer
Reference Manual
September 2019
00809-0100-5800
66
Emerson.com/Rosemount
Summary of Contents for Rosemount CT5800
Page 20: ...Contents Reference Manual September 2019 00809 0100 5800 xx Emerson com Rosemount ...
Page 34: ...Plan Reference Manual September 2019 00809 0100 5800 34 Emerson com Rosemount ...
Page 60: ...Startup procedure Reference Manual September 2019 00809 0100 5800 60 Emerson com Rosemount ...
Page 90: ...Gas calibration Reference Manual September 2019 00809 0100 5800 90 Emerson com Rosemount ...
Page 168: ...Shutdown procedure Reference Manual September 2019 00809 0100 5800 168 Emerson com Rosemount ...
Page 178: ...Theory of operation Reference Manual September 2019 00809 0100 5800 178 Emerson com Rosemount ...
Page 180: ...Spare parts list Reference Manual September 2019 00809 0100 5800 180 Emerson com Rosemount ...