The user shall ensure that the flow of process gas is limited to a maximum flow rate of
6 liters per minute.
•
The equipment shall only be used with process gases which are in gas groups B, C, or D
(Divisions) or IIB + H
2
(Zones) and must not contain oxygen or any other oxidizer in
concentrations greater than those found in normal air.
•
This assessment does not cover reliable function, performance, or other properties of
the equipment not related to safety.
•
The equipment is to be installed using wire no larger than 1 mm
2
(18 AWG).
•
The equipment is only to be installed by manufacturer trained personnel.
•
If at any time there is a conflict between the system safety provisions and any relevant
local (national or regional) requirements, the local requirements always take
precedence.
•
The equipment is not to be used with flammable liquids.
Reference Manual
ATEX / IECEx Assessment Report Summary
00809-0100-5800
September 2019
Rosemount CT5800
175
Summary of Contents for Rosemount CT5800
Page 20: ...Contents Reference Manual September 2019 00809 0100 5800 xx Emerson com Rosemount ...
Page 34: ...Plan Reference Manual September 2019 00809 0100 5800 34 Emerson com Rosemount ...
Page 60: ...Startup procedure Reference Manual September 2019 00809 0100 5800 60 Emerson com Rosemount ...
Page 90: ...Gas calibration Reference Manual September 2019 00809 0100 5800 90 Emerson com Rosemount ...
Page 168: ...Shutdown procedure Reference Manual September 2019 00809 0100 5800 168 Emerson com Rosemount ...
Page 178: ...Theory of operation Reference Manual September 2019 00809 0100 5800 178 Emerson com Rosemount ...
Page 180: ...Spare parts list Reference Manual September 2019 00809 0100 5800 180 Emerson com Rosemount ...