WARNING
EXPLOSION HAZARD
This article is in accordance with IEC 60079-0: 2011 Clause 30.
Failure to observe this warning and/or follow safety instructions could result in an
explosion or potentially hazardous situaiton which, if not avoided, could result in death or
serious injury.
Comply with all relevant national legislative requirements and regulations when
installing and wiring this analyzer.
Consider all safety instructions within this manual and all associated instruction
manuals.
WARNING
EXPLOSION HAZARD
Installing the instrument requires opening the enclosure and working at the open
instrument. This is permitted only when both no hazardous atmosphere is present and
the instrument and connected external circuitry are de-energized.
Depending on the local regulations, this may require a competent hot work supervisor
to issue a hot work permit.
CAUTION
SHOCK AND VIBRATION
Failure to follow this caution may result in damage to the analyzer.
The Rosemount CT5800 contains sensitive electronic equipment.
Do not subject it to any shock or vibration.
Procedure
1. On receipt of goods, look for any visible damage to the analyzer and verify that all
items noted to be shipped were received. Record on the goods receipt note any
damage or missing items, noting both the item(s) and the quantity missing.
2. Visually inspect the exterior of the analyzer for signs of damage, corrosion, gas
leaks, or signs of previously overheating.
3. Report anything found to the maintenance organization.
4. Attach suitably rated and tested lifting slings to the safety engineered lifting eye
bolts mounted on the sides of the enclosure.
5. One person carefully guides the equipment from the horizontal to the vertical
position while the other person lifts the equipment.
6. Use safety approved and tested lifting equipment to remove the analyzer from the
shipping container.
7. Place the analyzer on a solid, level surface and prepare to wall mount the analyzer.
8. Make sure that the analyzer is stored in its protective plastic cover until installation.
Reference Manual
Plan
00809-0100-5800
September 2019
Rosemount CT5800
33
Summary of Contents for Rosemount CT5800
Page 20: ...Contents Reference Manual September 2019 00809 0100 5800 xx Emerson com Rosemount ...
Page 34: ...Plan Reference Manual September 2019 00809 0100 5800 34 Emerson com Rosemount ...
Page 60: ...Startup procedure Reference Manual September 2019 00809 0100 5800 60 Emerson com Rosemount ...
Page 90: ...Gas calibration Reference Manual September 2019 00809 0100 5800 90 Emerson com Rosemount ...
Page 168: ...Shutdown procedure Reference Manual September 2019 00809 0100 5800 168 Emerson com Rosemount ...
Page 178: ...Theory of operation Reference Manual September 2019 00809 0100 5800 178 Emerson com Rosemount ...
Page 180: ...Spare parts list Reference Manual September 2019 00809 0100 5800 180 Emerson com Rosemount ...