Figure 7-33: Replacing a Temperature Controller
A. Temperature controller
Procedure
1. Shut down the analyzer as described in
.
2. Open the enclosure as described in
3. Press the small lever. While holding the lever in the down position, unplug the
temperature controller (A) from the upper DIN rail.
NOTICE
When replacing a temperature controller, the interconnecting wiring does not have
to be disconnected from the base unit.
4. Discard the unserviceable temperature controller.
5. Inspect the replacement temperature controller for damage. If damage has
occurred, contact your local Customer Care Representative.
6. Fit the replacement temperature controller (A) onto the upper DIN rail. Make sure
the orientation of the temperature controller is correct. Press the small lever (C) on
the base unit and, while holding the lever in the down position, push the
temperature controller home into the base unit. Release the lever and check that
the temperature controller is secure.
7. Attach the cables and wiring connections to the temperature controller.
8. Remove the protective masking tape from the flanges, seal, and flamepath.
Reference Manual
Troubleshooting and diagnostics
00809-0100-5800
September 2019
Rosemount CT5800
139
Summary of Contents for Rosemount CT5800
Page 20: ...Contents Reference Manual September 2019 00809 0100 5800 xx Emerson com Rosemount ...
Page 34: ...Plan Reference Manual September 2019 00809 0100 5800 34 Emerson com Rosemount ...
Page 60: ...Startup procedure Reference Manual September 2019 00809 0100 5800 60 Emerson com Rosemount ...
Page 90: ...Gas calibration Reference Manual September 2019 00809 0100 5800 90 Emerson com Rosemount ...
Page 168: ...Shutdown procedure Reference Manual September 2019 00809 0100 5800 168 Emerson com Rosemount ...
Page 178: ...Theory of operation Reference Manual September 2019 00809 0100 5800 178 Emerson com Rosemount ...
Page 180: ...Spare parts list Reference Manual September 2019 00809 0100 5800 180 Emerson com Rosemount ...