WARNING
BURNS
Some parts of the analyzer may be heated to 150 °F (65 °C). Failure to allow sufficient
cooling may cause serious burn injury to personnel.
Do not touch any part of the analyzer.
Assume all parts of the analyzer are hot unless it has been switched off and allowed to
cool down.
Always wear proper protective equipment when handling the analyzer.
Switch off the analyzer and allow it to cool for at least two hours before fitting,
removing, or performing any maintenance.
The analysis cell is insulated against heat loss. Allow the analyzer to cool for at least 12
hours before performing any maintenance on, or in the vicinity of, the analysis cell.
If burns occur, seek immediate medical treatment.
WARNING
HEAVY ITEM
The analyzer weighs 176.4 lb. (80 kg). Failure to properly handle the analyzer may cause
injury to personnel.
Use a minimum of two people when lifting and transporting the equipment.
Wear suitable protective gloves and protective footwear.
NOTICE
As a general principle, if any optical component other than the cell assembly and the laser
modules is unserviceable, then the manufacturer must repair the analyzer. This is because
the repair, replacement, and alignment of the optical components requires the use of
special optical test/calibration equipment and procedures.
7.8
Tools required for troubleshooting
The following tools are required to remove and replace components
Table 7-1: Required Tools
14 mm
Hex key needed to open the box
2.5 mm
Hex Socket (Allan key)
4.0 mm
Hex Socket (Allan key)
5.0 mm
Hex Socket (Allan key)
7.0 mm
Spanner
8.0 mm
Spanner
14 mm
Spanner
16 mm
Spanner
Reference Manual
Troubleshooting and diagnostics
00809-0100-5800
September 2019
Rosemount CT5800
109
Summary of Contents for Rosemount CT5800
Page 20: ...Contents Reference Manual September 2019 00809 0100 5800 xx Emerson com Rosemount ...
Page 34: ...Plan Reference Manual September 2019 00809 0100 5800 34 Emerson com Rosemount ...
Page 60: ...Startup procedure Reference Manual September 2019 00809 0100 5800 60 Emerson com Rosemount ...
Page 90: ...Gas calibration Reference Manual September 2019 00809 0100 5800 90 Emerson com Rosemount ...
Page 168: ...Shutdown procedure Reference Manual September 2019 00809 0100 5800 168 Emerson com Rosemount ...
Page 178: ...Theory of operation Reference Manual September 2019 00809 0100 5800 178 Emerson com Rosemount ...
Page 180: ...Spare parts list Reference Manual September 2019 00809 0100 5800 180 Emerson com Rosemount ...