Figure 7-44: Gas Input and Output Lines
A. Gas input and output lines
4. Remove the two M5 socket head cap screws, M5 Spring washers, and M5 plain
washers securing the DIN rail support bracket to the base of the enclosure.
Retain the screws and washers for future use.
5. Pull the DIN rail away from the enclosure.
6. Unscrew the two 14 mm connections to the gas input and output lines.
See
7. Replace the gas input and output lines as required.
7.16.4
Replacing the analysis cell
DANGER
ELECTRIC SHOCK
This article is in accordance with IEC 60079-0: 2011 Clause 30.
Death or personal injury may result if this is not observed.
The analyzer uses mains voltage that is dangerous to life. Make sure that the circuit
breakers are set to OFF and tagged off before opening the front cover.
Reference Manual
Troubleshooting and diagnostics
00809-0100-5800
September 2019
Rosemount CT5800
157