5
Verifying gas concentrations
5.1
Verification
Verification flows the known gas concentration through the analyzer and gives you a
display of the measurement, the cylinder value, and the difference between the two. You
can use verification to confirm that the analyzer is within tolerance. If it is out of tolerance,
perform a calibration (see
).
5.1.1
Zero verification
Zero verification confirms that when no sample gas is flowing through the analyzer, the
gas concentrations measured by the instrument are zero. To zero verify, compare the
analyzer measurements to a known sample gas using the following procedure:
Procedure
1. Refer to
. Using
and
, select
GAS SERVICE
.
2. Press
.
The
Select gas
screen (
Figure 5-1: Select gas Screen
3. Use
and
to move the cursor until the gas that you wish to zero verify is
selected. Press
.
The
Select Type
Reference Manual
Verifying gas concentrations
00809-0100-5800
September 2019
Rosemount CT5800
71
Summary of Contents for Rosemount CT5800
Page 20: ...Contents Reference Manual September 2019 00809 0100 5800 xx Emerson com Rosemount ...
Page 34: ...Plan Reference Manual September 2019 00809 0100 5800 34 Emerson com Rosemount ...
Page 60: ...Startup procedure Reference Manual September 2019 00809 0100 5800 60 Emerson com Rosemount ...
Page 90: ...Gas calibration Reference Manual September 2019 00809 0100 5800 90 Emerson com Rosemount ...
Page 168: ...Shutdown procedure Reference Manual September 2019 00809 0100 5800 168 Emerson com Rosemount ...
Page 178: ...Theory of operation Reference Manual September 2019 00809 0100 5800 178 Emerson com Rosemount ...
Page 180: ...Spare parts list Reference Manual September 2019 00809 0100 5800 180 Emerson com Rosemount ...