Table 2-2: Power, Digital, and Analog Connections
(continued)
Terminal
Function
24
Status output 2 (maintenance required)
25
Status output 3 (out of specification)
26
Status output 4 (failed)
Table 2-3: Ethernet Interface Module wiring to MOXA Ethernet Socket
Terminal
Designation
Color coding
1
Tx+
Orange / White
2
Tx
Orange
3
Rx+
Green / White
4
Not used
Blue
5
Not used
Blue / White
6
Rx-
Green
7
Not used
Brown / White
8
Not used
Brown
2.3.5
Connecting the sample supply and return lines
Sample gas supply and sample return connections are from
¼
-in. (6 mm) Swagelok
®
type
fittings.
The analyzer has one gas input and one gas output, which are located on the base of the
analyzer (
).
The gas sample that is to be measured for impurities enters the instrument through the
sample gas input port.
Once the sample gas has been examined for impurities, it is expelled from the instrument
through the sample gas return port.
NOTICE
Remove the breather blanking cap and never connect anything to this port.
To avoid the risk of gas leaks, make sure that these connections are made correctly and
tightly. Both the sample gas supply pipe and the sample return pipe should be thermal
insulated.
NOTICE
The Swagelok recommendation for pipe fittings of this size is to tighten the nut finger
tight and then tighten an additional one and a quarter (1 and
¼
) turns with a spanner.
The maximum gas supply pressure is 1 BarG.
Reference Manual
Install
00809-0100-5800
September 2019
Rosemount CT5800
49
Summary of Contents for Rosemount CT5800
Page 20: ...Contents Reference Manual September 2019 00809 0100 5800 xx Emerson com Rosemount ...
Page 34: ...Plan Reference Manual September 2019 00809 0100 5800 34 Emerson com Rosemount ...
Page 60: ...Startup procedure Reference Manual September 2019 00809 0100 5800 60 Emerson com Rosemount ...
Page 90: ...Gas calibration Reference Manual September 2019 00809 0100 5800 90 Emerson com Rosemount ...
Page 168: ...Shutdown procedure Reference Manual September 2019 00809 0100 5800 168 Emerson com Rosemount ...
Page 178: ...Theory of operation Reference Manual September 2019 00809 0100 5800 178 Emerson com Rosemount ...
Page 180: ...Spare parts list Reference Manual September 2019 00809 0100 5800 180 Emerson com Rosemount ...