Figure 1-5: Rosemount CT5800 Radius for Opening
A. Opening arc
Table 1-4: Environmental Characteristics
Environmental characteristic
Value
Comment
Operating temperature range
-4 to 131 °F (-20 to 55 °C)
Ambient temperature
Sample gas temperature range
(condition for safe use Europe)
39 to 140 °F (4 to 60 °C)
Factory set, specify on order
Sample gas temperature range
(condition for safe use North America)
122 to 176 °F (50 to 80 °C)
Factory set, specify on order
Sample gas moisture content
8%
Maximum
Sample gas particulate density
5 mg/m
3
Maximum
Sample gas particulate size
10 µm
Maximum
IP code
66
IP to IEC 60529
Sensor humidity range
10 to 95%
Relative humidity (non-condensing) at
113 °F (45 °C)
1.9.1
Optical description
The laser modules are located in the core of the analyzer. Each laser module produces a
separate light beam, and these beams are combined linearly as the modules are aligned in
Reference Manual
Plan
00809-0100-5800
September 2019
Rosemount CT5800
31
Summary of Contents for Rosemount CT5800
Page 20: ...Contents Reference Manual September 2019 00809 0100 5800 xx Emerson com Rosemount ...
Page 34: ...Plan Reference Manual September 2019 00809 0100 5800 34 Emerson com Rosemount ...
Page 60: ...Startup procedure Reference Manual September 2019 00809 0100 5800 60 Emerson com Rosemount ...
Page 90: ...Gas calibration Reference Manual September 2019 00809 0100 5800 90 Emerson com Rosemount ...
Page 168: ...Shutdown procedure Reference Manual September 2019 00809 0100 5800 168 Emerson com Rosemount ...
Page 178: ...Theory of operation Reference Manual September 2019 00809 0100 5800 178 Emerson com Rosemount ...
Page 180: ...Spare parts list Reference Manual September 2019 00809 0100 5800 180 Emerson com Rosemount ...