2.3.4
Connecting the signal cables
Connect the signal cables through conduit outlets 2 and 3 as shown in the figures below.
NOTICE
This section is in accordance with IEC 60079-0: 2011 Clause 30. This section must not be
changed, amended, or removed.
The signal cables are connected to the system through conduit outlets 2 and 3 as shown in
. All signal cables are to be minimum 20 AWG tri-rated
switchgear cables. Customer supplied conduit and cables must be terminated in the
conduit outlets in accordance with local electrical codes.
Figure 2-7: Cable Glands, Conduits, and Gas Line Connections
A. Breather
B. Gas connection out (flame arrestor) (M18 x 1.5)
C. Gas connection in (flame arrestor) (M18 x 1.5)
D. Conduit / cable entry apertures (M20 x 1.5)
Figure 2-8: Signal connections
Reference Manual
Install
00809-0100-5800
September 2019
Rosemount CT5800
47