Figure 5-4: Verify Zero Screen
7. Allow the concentrations to stabilize and wait for two minutes after stabilization.
This screen gives a reading of the concentration of the selected gas that is present
as an impurity in the nitrogen calibration gas, as measured by the instrument. In the
example (
), the gas being measured is NH
3
(ammonia), and the
instrument has detected a concentration of 0.40 ppm.
8. If the reading is within tolerance, no further action is required. Press
to end the
zero verification process.
The display controller proceeds to the
Result
screen (
Figure 5-5: Result Screen
9. If the reading is outside tolerance, you need to zero calibrate the analyzer.
10. In either case, press
to return to the
Calibration / Verification Complete
screen
).
Reference Manual
Verifying gas concentrations
00809-0100-5800
September 2019
Rosemount CT5800
73