Figure 7-36: Installing the Lower DIN Rail
3. Connect the two 14 mm gas input and output line connectors at the flame arrestor.
Replacing the gas input and output lines
4. Connect all cables and wiring connections as appropriate.
5. Remove the protective masking tape from the flanges, seal, and flamepath.
CAUTION
PROTECT FLAMEPATH
Any damage to the flamepath will invalidate certification.
Use care not to damage the flamepath.
6. Close the enclosure as described in
and secure with the
captive M16 bolts and engage the threads on the rear housing.
7. Apply power to the analyzer.