9. Visually check that the vent cap is fitted to the purge vent (
Figure 1-4
).
UNDER NO
CIRCUMSTANCES
must the analyzer be started with the protector cap fitted as this
will prevent the analyzer from venting during the purge process.
WARNING
FLAMMABLE/EXPLOSIVE GASES
The purge/pressurization cycle expels any flammable or explosive gases from the
electrical compartment of the . Wait until the purge/pressurization cycle is
complete before applying electrical power to the rest of the analyzer.
10. Turn on the compressed air supply to the Rosemount CT5100(Ex).
11. Set the main circuit breaker to
On
. The air pressurization will begin its automatic
start-up and begin to purge the air inside the Rosemount CT5100(Ex).
Figure 4-3: Purge Control Panel
A. K1 LED
B. P/SV LED
C. PT100 LED
12. Observe the PT100 LED on the purge control panel (
Figure 4-3
).
If the PT100 LED (C) illuminates red, the air pressurization system is unserviceable,
and the start-up procedure must be halted immediately. To halt the start-up
procedure, set the external main circuit breaker to
Off
. lock off or tag off the main
Quick Start Guide
Start-up procedure
D-7010-0055
August 2018
Quick Start Guide
49
Summary of Contents for Rosemount CT5100 Ex
Page 1: ...Quick Start Guide D 7010 0055 Rev C August 2018 Rosemount CT5100 Ex Continuous Gas Analyzer ...
Page 12: ...Contents Quick Start Guide August 2018 D 7010 0055 12 Rosemount CT5100 Ex ...
Page 34: ...Install Quick Start Guide August 2018 D 7010 0055 34 Rosemount CT5100 Ex ...
Page 44: ...Operating the analyzer Quick Start Guide August 2018 D 7010 0055 44 Rosemount CT5100 Ex ...
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......