Once complete the program will automatically open the
Result
screen (
Figure 5-15
).
Figure 5-15: Result Screen
7. Press
OK
.
The
Verify next
screen displays.
Figure 5-16: Verify Next Step Screen
8. Press
DONE
to finish.
•
To calibrate additional gases, select
GAS
and repeat
Step 1
through
Step 8
.
•
To change types select
TYPE
.
This completes the process and the display will return to the
Main
menu.
5.2.2
To span calibrate an individual gas
Procedure
1. To get the
Main menu
, press the
MENU
button on the gas sensor main screen,
Figure 5-2
or
Figure 5-3
.
Quick Start Guide
Calibrating the analyzer
D-7010-0055
August 2018
Quick Start Guide
67