WARNING
TRANSPORTATION HAZARD
Use safety approved lifting equipment. The user must ensure the equipment is tested,
meets the lifting ratings for the weight of the equipment, and is in good operational
condition.
Failure to verify equipment meets the lifting ratings and is in good operational condition
may cause injury to personnel or damage the equipment.
Procedure
1. Visually inspect the exterior of the analyzer for signs of damage, corrosion, gas
leaks, or signs of previously overheating.
2. Report anything found to the maintenance organization.
3. Attach suitably rated and test lifting slings to the safety engineered lifting eye bolts
mounted on top of the analyzer.
4. One person should carefully guide the equipment from the horizontal to vertical
position while the other person lifts the equipment.
5. Use safety approved and tested lifting equipment to remove the analyzer from the
shipping container.