NOTICE
Pressing the
BACK
button will cancel the procedure and return the user to the
previous screen.
2. Use the middle buttons ( and ) to scroll up or down to select
GAS SERVICE
.
Press the
SEL
button.
Figure 5-25: Main Menu (GAS SERVICE)
3. Use the middle buttons ( and ) to scroll up or down to select
CALIBR ALL
.
Press the
SEL
button.
Figure 5-26: Gas Service (CALIBR ALL)
4. The system will work through the
ZERO
and
SPAN
calibration process for all
readings.
The
Please wait
screen will be displayed. This takes approximately one minute per
gas depending on the purge time.
Calibrating the analyzer
Quick Start Guide
August 2018
D-7010-0055
72
Rosemount
™
CT5100 Ex