Figure 5-8: Input Type
Figure 5-9: Calibration
13. Using the
LEFT, RIGHT, INCR
(increase) and
DECR
(decrease) buttons to enter the
desired offset (in the example above, 1.06ppm).
14. Press
OK
to confirm.
The factory default span factor is one, and under normal circumstances, the offset
should be between 0.8 and 1.2.
5.2.1
To zero calibrate an individual gas
Procedure
1. To get the
Main menu
, press the
MENU
button on the gas sensor main screen,
Figure 5-2
or
Figure 5-3
.
NOTICE
Pressing the
BACK
button will cancel the procedure and return the user to the
previous screen.
2. Use the middle buttons ( and ) to scroll up or down to select
GAS SERVICE
.
Calibrating the analyzer
Quick Start Guide
August 2018
D-7010-0055
64
Rosemount
™
CT5100 Ex
Summary of Contents for Rosemount CT5100 Ex