Figure 5-27: Calibration
On completion, the
Finished
screen (
Figure 5-28
) will display.
Figure 5-28: Calibration (finished)
5. Press
BACK
.
This completes the process and the display will return to the
Main menu
.
5.2.4
To change the automatic calibration settings
Procedure
1. To get the
Main menu
, press the
MENU
button on the gas sensor main screen,
Figure 5-2
or
Figure 5-3
.
NOTICE
Pressing the
BACK
button will cancel the procedure and return the user to the
previous screen.
2. Use the middle buttons ( and ) to scroll up or down to select
GAS SERVICE
.
Press the
SEL
button.
Quick Start Guide
Calibrating the analyzer
D-7010-0055
August 2018
Quick Start Guide
73