WARNING
EXPLOSION HAZARD
In accordance with IEC 60079-0: 2011 Clause 30.
The start-up procedure is only to be carried out by properly trained personnel who understand the contents of all applicable
manuals and related instructions.
Failure to observe this warning could cause an explosion or potentially hazardous situation, which if not avoided, may cause death,
personal injury, and/or damage to persons and/or property.
WARNING
EXPLOSION HAZARD
In accordance with IEC 60079-0: 2011 Clause 30.
During the pre-purge phase all inputs and outputs connected to external equipment MUST be shut off.
This will ensure that no hazardous voltages are present within the analyzer enclosure when not pressurized.
Note: the internal backup battery is still connected and associated circuitry remains powered.
Failure to observe this warning could cause an explosion or potentially hazardous situation, which if not avoided, may cause death,
personal injury, and/or damage to persons and/or property.
WARNING
EXPLOSION HAZARD
In accordance with IEC 60079-0: 2011 Clause 30.
Do not open while an explosive atmosphere may be present.
Failure to observe this warning could cause an explosion or potentially hazardous situation, which if not avoided, may cause death,
personal injury, and/or damage to persons and/or property.
WARNING
EXPLOSION HAZARD
In accordance with IEC 60079-0: 2011 Clause 30.
Do not keep operating the analyzer if the enclosure shows permanent deformations after performing the overpressure test.
Failure to observe this warning could cause an explosion or potentially hazardous situation, which if not avoided, may cause death,
personal injury, and/or damage to persons and/or property.
WARNING
EXPLOSION HAZARD
In accordance with IEC 60079-0: 2011 Clause 30.
Use only replacement parts and components authorized by Emerson.
All replacement parts and components must be certified and approved for use in hazardous areas.
Failure to comply will void certification and may cause an explosion or potentially hazardous situation, which if not avoided, may
cause death, personal injury, and/or damage to persons and/or property.
8
Summary of Contents for Rosemount CT5100 Ex
Page 1: ...Quick Start Guide D 7010 0055 Rev C August 2018 Rosemount CT5100 Ex Continuous Gas Analyzer ...
Page 12: ...Contents Quick Start Guide August 2018 D 7010 0055 12 Rosemount CT5100 Ex ...
Page 34: ...Install Quick Start Guide August 2018 D 7010 0055 34 Rosemount CT5100 Ex ...
Page 44: ...Operating the analyzer Quick Start Guide August 2018 D 7010 0055 44 Rosemount CT5100 Ex ...
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......