Figure 5-23: Result
10. Press
DONE
to finish. Refer to
Figure 5-24
.
Figure 5-24: Calibration Complete
This completes the process and the display will return to the
Main menu
.
5.2.3
To calibrate all gases (zero and span)
Prerequisites
On systems with valve control, it is possible to run a calibration on all gases. The valves will
sequentially switch through each of the span and zero gases. This option is not practical to
use on systems which require manual valve switching.
In order to set the cylinder concentration when a new span gas cylinder is to be used, it is
necessary to perform a single gas calibration as described earlier. This concentration will
then be stored for future calibrations.
Procedure
1. To get the
Main menu
, press the
MENU
button on the gas sensor main screen,
Figure 5-2
or
Figure 5-3
.
Quick Start Guide
Calibrating the analyzer
D-7010-0055
August 2018
Quick Start Guide
71