CAUTION
Under no circumstance
MUST
the top two buttons on the display controller be pressed
simultaneously. This will override the system code supplied with the analyzer and display
the default program screen for the controller.
Figure 4-7: Default Program Screen
This will lock the software controlling the Rosemount CT5100(Ex) requiring the analyzer to
be shutdown externally and restarted risking a possible corruption of the software and loss
of the analyzer operation during the shutdown and restart process.
It is essential that only trained, qualified personnel operate the controls on the analyzer.
Failure to follow these instructions may cause warranty invalidation, property damage
and/or personal injury or death.
The purge control panel (
Figure 4-8
) is used to program the operation of the overpressure
air system that prevents flammable and explosive gases from entering the electrical
compartment of the Rosemount CT5100(Ex). Adjustment of the overpressure air system
must only be performed by maintenance personnel.
Five LEDs, identified as K1, K2, PSV, Bypass and PT100 are located on the upper part of the
purge control panel. These five LEDs indicate the status of the overpressure air system.
Quick Start Guide
Start-up procedure
D-7010-0055
August 2018
Quick Start Guide
53
Summary of Contents for Rosemount CT5100 Ex
Page 1: ...Quick Start Guide D 7010 0055 Rev C August 2018 Rosemount CT5100 Ex Continuous Gas Analyzer ...
Page 12: ...Contents Quick Start Guide August 2018 D 7010 0055 12 Rosemount CT5100 Ex ...
Page 34: ...Install Quick Start Guide August 2018 D 7010 0055 34 Rosemount CT5100 Ex ...
Page 44: ...Operating the analyzer Quick Start Guide August 2018 D 7010 0055 44 Rosemount CT5100 Ex ...
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......