DL8000 Preset Instruction Manual
Revised February-2016
DL8000 Keypad Display
E-61
Note
:
Whether you select single and multiple meter factors changes the
process of product calibration slightly. Differences, as
appropriate, are noted.
Select
PRODUCT Calib
to start the product meter calibration process.
14/03/12 12:26:02
For traversing use
UP | DOWN
LEFT | RIGHT
ENTER to continue
EXIT to Go Back
EXIT to quit
Additive Calib
Analog Calib
RTD Calib
PRODUCT Calib
(Pg:1/1)
The system then displays a Prover Selection menu, showing the
supported proving methods.
Note
:
The menu displays Can Proving, which is the
only
proving
method currently supported.
14/03/12 12:26:02
Additive Calib
Analog Calib
RTD Calib
PRODUCT Calib.
Can Prover
Pg:1/1)
Press
ENTER
to continue or
EXIT
to halt the calibration process and
return to the Calibration menu.
You must check the following preconditions before proceeding:
Check Transaction status
Check WM status
Check
Transaction Status
Check the transaction status first from the Status Flag [63,0,119]. If you
select product calibration and a transaction has not ended, the system
displays the following message:
14/03/12 12:26:02
EXIT to Quit
Transaction Inprog
Stop To End Trans