DL8000 Preset Instruction Manual
Revised February-2016
DL8000 Keypad Display
E-57
After you provide all the parameters required for calibration press
ENTER
to continue. The program then displays the live value from an
external source with a corresponding “dead” (expected) value you can
change.
14/03/12 12:26:02
Enter ZERO EU
Enter SPAN EU
(ENTER to update )
(STOP to End )
Live Dead
80.3000003 80.5
(0-9, .)
Press
ENTER
to update the value. You then enter a span value:
14/03/12 12:26:02
Enter ZERO EU
Enter SPAN EU
(ENTER to update )
(STOP to End )
Live Dead
80.3000003 80.5
229.000000 228.99
(0-9, .)
If you need mid-points for calibration, enter those next.
The program checks all values you enter. If any value is not within
accepted ranges, the program displays the error message
InvdEntry!ReEnter
:
14/03/12 12:26:02
Enter ZERO EU
Enter SPAN EU
(ENTER to update )
(STOP to End )
Live Dead
80.3000003 80.5
229.000000 228.99
InvdEntry!ReEnter
(0-9, .)
You must re-enter those values in order to proceed.
Valid ranges are:
RTD unit
Minimum Range Maximum Range
Deg F
-58.0
662.0
Deg C
-50.0
350.0
Deg K
223.15
623.15
Deg R
401.67
1121.67
OHM
80.31
229.72