18
2. SAFETY REGULATIONS
Full-electronic G1-1 HV DC-Tester
the device indicates the fault message „Attention residual voltage“.
Note:
The residual voltage measurement can
not
switched off on the
„Setup“ menu or using an interface command.
Automatic discharging is performed only on DC test devices. At AC test
devices the high voltage is switched off on zero-crossing.
Attention:
Leave the test object connected until it is discharged. The
device indicates this state with the illumination of the green RUN button.
Additional discharging via an external resistor
The device has an output for the connection of an external discharge re-
sistor. At the end of the HV/ IS test this output is connected by the device.
On this topic see chapter 6.2.5.1 „Outputs/external discharge relay“.