11. TEST PARAMETER
Full-electronic G1-1 HV DC-Tester
tripping current (mA)
minimum current (mA)
0.05
1.50
HV_1000V
Tripping current (0.01 ... 2.0/5.0 mA)
: - the current entered is the maxi-
mum current that is allowed to flow during the HV testing. If this value
is exceeded, an HV fault (I>Imax) is indicated (test object fault - HV
flashover). The tripping current (Imax) is dependent on the measure
-
ment range.
Minimum current (0.00 ... dynamic)
: - the current entered must flow dur
-
ing the HV testing as a minimum. If the current is less than this value,
a contact fault (I<Imin) is indicated.
off
- no minimum voltage monitoring.
on
-
(0.00 up to tripping current )
minimum current monitoring
active. The current range is all times 0.1 / 0.01 / 0.001 lower
then the tripping current.
Fig. 11.1d: HV test parameter 3