E-8
WHAT AND HOW TO CHECK
REMEDIAL ACTION
3-1.
(1)
Check
or
switch of X-axis; the procedure
is the same as 1-4 (1).
3-1.
(1)
Same as 1-4 (1).
(2)
Check voltage at XL or XR terminal of PB6
(EAUA--087703) with a tester (synchroscope).
(2)
If it is approx. 0V, the limit switch is normal, and
therefore check other parts. If it is approx. 5V, repair
(replace)
the
limit
switch.
(Also
check
the
connecting cable.)
(3)
If no fault is found in item (2), replace the board
(PB6) with a spare one.
(3)
If there is no change, check other parts. If it operates
correctly, the board is defective.
(4)
Check the operation after replacing PB7 (EAUA--
087800) with a spare P.C. board.
(4)
Same as in (3). (Refer to Section 6 for checking
signals.)
(5)
Check the operation after replacing PB9 (EAUA--
023600) with a spare P.C. board.
(5)
Same as in (3). (Refer to Section 7 for checking
signals.)
(6)
Check the cable connecting MF4 and Motor.
(6)
If the cable is open circuit, repair it.
3-2.
(1)
Check the operation replacing the panel P.C. board.
3-2.
(1)
If there is no change, check other parts.
(Also see items (2) and (3).)
(2)
Check the operation after replacing PB6 (EAUA--
087703) with spare P.C. board.
(2)
Same as in (1).
(3)
Check the operation after replacing PB7 (EAUA--
087800) with spare P.C. board.
(3)
Same as in (1).
3-3.
(1)
Same as in 3-1 (2).
3-3.
(1)
If it is approx. +5V at the limit position, see item (2).
If it is approx. 0V, repair (replace) the limit switch.
(Also check connecting cable.)
(2)
Same as in 3-1 (3).
(2)
Same as in 3-1 (3).
3-4.
(1)
Check the DATA contents.
3-4.
(1)
If the data is correct, check other points. If the data is
incorrect, re-enter data. (When the data is not stored,
refer to 2-4.)
(2)
Check
and
switches of Y-axis: The
procedure is the same as in 3-1. (Terminals to be
checked are YF and FR.)
(2)
Same as in 3-1.
3-5.
(1)
Check
and
switches of Z-axis: The
procedure is the same as in 3-1. (Terminals to be
checked are ZU and ZD)
3-5.
(1)
Same as in 3-1.
891110
Summary of Contents for DAD 2H/6T
Page 5: ...INTRODUCTION...
Page 7: ...A GENERAL...
Page 9: ...B CONSTRUCTION...
Page 11: ...B 2 2 Composition and Arrangement of Electrical Equipment Fig B 2 910810 910810...
Page 12: ...B 3 891110 3 Pneumatic Water Schematic Diagram Fig B 3 891110...
Page 13: ...C CIRCUIT DIAGRAM...
Page 16: ......
Page 17: ......
Page 18: ......
Page 20: ......
Page 21: ......
Page 22: ......
Page 23: ......
Page 24: ......
Page 25: ......
Page 26: ......
Page 28: ......
Page 30: ......
Page 32: ......
Page 33: ......
Page 34: ......
Page 35: ......
Page 36: ......
Page 37: ......
Page 38: ......
Page 39: ......
Page 40: ......
Page 41: ......
Page 42: ......
Page 44: ......
Page 46: ......
Page 48: ......
Page 50: ......
Page 51: ......
Page 52: ......
Page 54: ......
Page 56: ......
Page 58: ......
Page 60: ......
Page 61: ......
Page 62: ......
Page 64: ......
Page 65: ......
Page 66: ......
Page 68: ......
Page 69: ......
Page 70: ......
Page 72: ...EAUA 087600...
Page 74: ...EAUA 087703...
Page 76: ...EAUA 087800 X Y EAUA 087900 Z...
Page 78: ...EAUA 023600...
Page 81: ...D MAINTENANCE...
Page 83: ...E CHECKING PROCEDURES FOR TROUBLESHOOTING...
Page 110: ...E 27 Fig E 3 Output Waveform of Driver P C Board 840920...
Page 117: ...F ILLUSTRATION AND PARTS LIST...
Page 118: ...EXPLODED VIEW OF MAIN BODY A F 1...
Page 124: ...EXPLODED VIEW OF MAIN BODY B F 2...
Page 128: ...EXPLODED VIEW OF X AXIS 90 THETA F 3A...
Page 134: ...EXPLODED VIEW OF X AXIS MANUAL THETA F 3B MODEL DAD 2H 6TM...
Page 137: ...MEMO...
Page 138: ...EXPLODED VIEW OF X AXIS 360 THETA F 3C...
Page 141: ...MEMO...
Page 142: ...EXPLODED VIEW OF X AXIS F 3D MODEL DAD 2H 6T...
Page 144: ...EXPLODED VIEW OF Y AXIS F 4...
Page 148: ...EXPLODED VIEW OF Z AXIS F 5...
Page 154: ...EXPLODED VIEW OF PIPING A F 6...
Page 156: ...EXPLODED VIEW OF PIPING B F 7...
Page 165: ...G REGULAR INSPECTIONS...
Page 169: ...H CAMERA HEAD MONITOR TV...
Page 174: ...H 5 External Operating Unit Fig H 2 910110...
Page 175: ...H 6 Camera Head Internal Adjusters VIDEO PWB Fig H 3 891110...
Page 176: ...H 7 Monitor Internal Adjuster PROCESS PWB 891110 Fig H 4...
Page 177: ...H 8 Internal Monitor Adjusters MONITOR PWB 891110 Fig H 5...
Page 178: ...H 9 Internal Monitor Adjuster POWER SUPPLY PWB Fig H 6 891110...
Page 179: ...J MICROSCOPE MODEL DS...