Internal Component Replacement (Inside Front Enclosure)
10-54
Service Manual
7.
Connect the AC power inlet wire assembly as follows (
a.
Connect the 6-pin connector to P112 on the Power PCBA. Make sure that the
connector is oriented as shown in
, and that it is fully seated.
b.
Connect the 3-pin connector to CON1 on the power supply. Make sure that
the connector latch engages.
Note:
If a cable tie previously secured the AC power inlet wire assembly to the
power supply mounting boss (see
), install a new cable tie at the same
position. The cable tie should cause the wires connected to CON1 to form an arch
approximately 4 cm (1.5 in) above the connector. Trim off the end of the cable tie.
8.
Reinstall the cables on the equipotential stud as follows (
a.
Place the AC power inlet ground cable terminal on the stud.
b.
Place a lock washer on top of the terminal.
c.
Place a nut on top of the lock washer. Tighten the nut to 3.2 to 3.8 N-m (28.3
to 33.6 lb-in).
d.
Place the power supply ground cable terminal on top of the nut.
e.
Place a lock washer on top of the terminal.
f.
Place a nut on top of the lock washer. Tighten the nut to 3.2 to 3.8 N-m (28.3
to 33.6 lb-in).
9.
Follow the procedure
Rejoining the Front and Rear Enclosures
10.
Perform the following tests:
a.
b.
Summary of Contents for WarmTouch
Page 1: ...Service Manual WarmTouch Convective Warming Unit...
Page 8: ...Page Left Intentionally Blank vi Service Manual...
Page 32: ...Page Left Intentionally Blank 1 18 Service Manual Introduction...
Page 42: ...Installation 3 4 Service Manual Figure 3 3 Power Cord Wrapped and Secured...
Page 45: ...Installing the Warming Unit Service Manual 3 7 Figure 3 5 Warming Unit Mounted on IV Pole...
Page 49: ...Installing the Warming Unit Service Manual 3 11 Figure 3 9 Warming Unit Mounted on Cart...
Page 50: ...Page Left Intentionally Blank 3 12 Service Manual Installation...
Page 70: ...Page Left Intentionally Blank 5 12 Service Manual Maintenance...
Page 84: ...Page Left Intentionally Blank 6 14 Service Manual Troubleshooting...
Page 105: ...Power Cord Replacement Service Manual 8 7 Figure 8 5 Power Cord Wrapped and Secured...
Page 118: ...Page Left Intentionally Blank 9 8 Service Manual Accessing the Inside of the Warming Unit...
Page 246: ...Enclosure Replacement 12 14 Service Manual Figure 12 10 Power Cord Wrapped and Secured...
Page 266: ...Page Left Intentionally Blank 12 34 Service Manual Enclosure Replacement...
Page 274: ...Page Left Intentionally Blank 13 8 Service Manual Product Specifications...
Page 313: ......