Performance Specifications
Service Manual
13-3
13.5
Performance Specifications
13.6
Product Compliance
Maximum Contact Surface Temperature
44.1°C (111.4°F)
Average Time for Contact Surface Temperature
to Rise from 23°C ±2°C to 37°C
6 minutes
Average Time for Temperature of Air Exiting
the Hose to Rise from 23°C ±2°C to 37°C
< 1 minute
Accuracy of Displayed Temperature
±1.0°C (air entering hose)
Automatic Temperature Stepdown
(Boost to High Temperature)
After 45 minutes of continuous use,
blower will step down from the Boost
to High setting.
Thermal Protection Threshold
Thermostat (internal):
49°C to 55°C (120°F to 131°F)
Average Alarm Level
56 dB @ 1 meter
Equipment Classification
IEC 60601-1:2005, EN 60601-1:2006
ANSI/AAMI ES60601-1:2005
CAN/CSA C22.2 No 60601-1:08
IEC 80601-2-35:2009
EN 80601-2-35:2010
Protection Type
Class I
Degree of Protection
Type BF - Applied part
Mode of Operation
Continuous
Electromagnetic Compatibility
IEC/EN 60601-1-2:2007, Class A Emissions,
Non Life-Support Equipment. IEC/EN 80601-2-35;
Hazard Free to 10 V/m.
Protection Against Ingress of Fluids
Ordinary
Summary of Contents for WarmTouch
Page 1: ...Service Manual WarmTouch Convective Warming Unit...
Page 8: ...Page Left Intentionally Blank vi Service Manual...
Page 32: ...Page Left Intentionally Blank 1 18 Service Manual Introduction...
Page 42: ...Installation 3 4 Service Manual Figure 3 3 Power Cord Wrapped and Secured...
Page 45: ...Installing the Warming Unit Service Manual 3 7 Figure 3 5 Warming Unit Mounted on IV Pole...
Page 49: ...Installing the Warming Unit Service Manual 3 11 Figure 3 9 Warming Unit Mounted on Cart...
Page 50: ...Page Left Intentionally Blank 3 12 Service Manual Installation...
Page 70: ...Page Left Intentionally Blank 5 12 Service Manual Maintenance...
Page 84: ...Page Left Intentionally Blank 6 14 Service Manual Troubleshooting...
Page 105: ...Power Cord Replacement Service Manual 8 7 Figure 8 5 Power Cord Wrapped and Secured...
Page 118: ...Page Left Intentionally Blank 9 8 Service Manual Accessing the Inside of the Warming Unit...
Page 246: ...Enclosure Replacement 12 14 Service Manual Figure 12 10 Power Cord Wrapped and Secured...
Page 266: ...Page Left Intentionally Blank 12 34 Service Manual Enclosure Replacement...
Page 274: ...Page Left Intentionally Blank 13 8 Service Manual Product Specifications...
Page 313: ......