Principles of Operation
Service Manual
6-7
Thermostat
The thermostat is located in the hose duct adapter near the base of the hose
(see
on page 11-10). It is connected to the power PCBA. The ther-
mostat provides thermal cut-out protection by opening at 49°C to 55°C (120°F
to 131°F), resulting in heater and fan shutdown.
Thermistor
The thermistor is located in the hose duct adapter just below the screen at the
base of the hose (see
page 11-7). It is connected to the power PCBA.
During operation, the temperature measured by the thermistor is monitored
by the A/D processor and relayed to the UI microprocessor for display on the
LCD. If the temperature does not remain within specification, the software
shuts down the heater and fan and generates an alarm.
Summary of Contents for WarmTouch
Page 1: ...Service Manual WarmTouch Convective Warming Unit...
Page 8: ...Page Left Intentionally Blank vi Service Manual...
Page 32: ...Page Left Intentionally Blank 1 18 Service Manual Introduction...
Page 42: ...Installation 3 4 Service Manual Figure 3 3 Power Cord Wrapped and Secured...
Page 45: ...Installing the Warming Unit Service Manual 3 7 Figure 3 5 Warming Unit Mounted on IV Pole...
Page 49: ...Installing the Warming Unit Service Manual 3 11 Figure 3 9 Warming Unit Mounted on Cart...
Page 50: ...Page Left Intentionally Blank 3 12 Service Manual Installation...
Page 70: ...Page Left Intentionally Blank 5 12 Service Manual Maintenance...
Page 84: ...Page Left Intentionally Blank 6 14 Service Manual Troubleshooting...
Page 105: ...Power Cord Replacement Service Manual 8 7 Figure 8 5 Power Cord Wrapped and Secured...
Page 118: ...Page Left Intentionally Blank 9 8 Service Manual Accessing the Inside of the Warming Unit...
Page 246: ...Enclosure Replacement 12 14 Service Manual Figure 12 10 Power Cord Wrapped and Secured...
Page 266: ...Page Left Intentionally Blank 12 34 Service Manual Enclosure Replacement...
Page 274: ...Page Left Intentionally Blank 13 8 Service Manual Product Specifications...
Page 313: ......