![background image](http://html1.mh-extra.com/html/covidien/warmtouch/warmtouch_service-manual_2673928061.webp)
Safety Checks
Service Manual
5-3
5.4
Safety Checks
5.4.1
Physical Inspection
Covidien recommends that the customer perform a visual check of the follow-
ing items before each use of the warming unit:
•
Equipment —
Inspect the warming unit and power cord for mechanical damage
or deterioration.
•
Labels —
Inspect the safety labels for legibility.
After performing any service on the warming unit, be sure to check these items
before returning the unit to operation.
Covidien also recommends checking the time remaining time before the filter
must be replaced before returning the unit to operation. See
on page 5-10 for instructions.
5.4.2
Temperature and Alarm Verification
Once a year, the following tests must be performed:
•
Temperature Accuracy Test —
Use this test to verify heater output tempera-
for instructions.
•
Thermostat Test —
Use this test to verify the alarm system. See
instructions.
These tests are also required after certain service procedures. Refer to the
instructions in each component replacement section for testing requirements.
5.4.3
Electrical Safety Tests
The following electrical safety tests must be performed after certain service
procedures:
•
Ground Bond Test —
See
for values.
•
Earth Leakage Current Test (Line Leakage) —
See
for values.
•
Enclosure Leakage Current Test (Line Leakage) —
See
for values.
Refer to the instructions in each component replacement section for testing
requirements.
Summary of Contents for WarmTouch
Page 1: ...Service Manual WarmTouch Convective Warming Unit...
Page 8: ...Page Left Intentionally Blank vi Service Manual...
Page 32: ...Page Left Intentionally Blank 1 18 Service Manual Introduction...
Page 42: ...Installation 3 4 Service Manual Figure 3 3 Power Cord Wrapped and Secured...
Page 45: ...Installing the Warming Unit Service Manual 3 7 Figure 3 5 Warming Unit Mounted on IV Pole...
Page 49: ...Installing the Warming Unit Service Manual 3 11 Figure 3 9 Warming Unit Mounted on Cart...
Page 50: ...Page Left Intentionally Blank 3 12 Service Manual Installation...
Page 70: ...Page Left Intentionally Blank 5 12 Service Manual Maintenance...
Page 84: ...Page Left Intentionally Blank 6 14 Service Manual Troubleshooting...
Page 105: ...Power Cord Replacement Service Manual 8 7 Figure 8 5 Power Cord Wrapped and Secured...
Page 118: ...Page Left Intentionally Blank 9 8 Service Manual Accessing the Inside of the Warming Unit...
Page 246: ...Enclosure Replacement 12 14 Service Manual Figure 12 10 Power Cord Wrapped and Secured...
Page 266: ...Page Left Intentionally Blank 12 34 Service Manual Enclosure Replacement...
Page 274: ...Page Left Intentionally Blank 13 8 Service Manual Product Specifications...
Page 313: ......