Performance Verification
7-14
Service Manual
7.6
Electrical Safety Tests
This section provides values for electrical safety tests required by the warming
unit. Technicians must be familiar with the standards applicable to their respec-
tive institution and country. Test equipment and its application must comply
with the applicable standards.
7.6.1
Ground Bond Test
Without power cord: ≤ 100 m Ω
With power cord: ≤ 200 m Ω
7.6.2
Earth Leakage Current Test (Line Leakage)
7.6.3
Enclosure Leakage Current Test (Line Leakage)
Table 7-1.
Earth Leakage Current Test
AC Polarity
Line Cord
Neutral Cord
Leakage Current
Normal
Closed
Closed
300 μA
Reversed
Closed
Closed
300 μA
Normal
Open
Closed
1000 μA
Normal
Closed
Open
1000 μA
Table 7-2.
Enclosure Leakage Current Test
AC Line Cord
Neutral Line Cord
Power Line
Ground Cable
AAMI / ANSI
Standard ES1
Normal
Closed
Closed
100 μA
Normal
Closed
Open
500 μA
Normal
Open
Closed
500 μA
Reversed
Closed
Closed
100 μA
Reversed
Open
Closed
500 μA
Reversed
Closed
Open
500 μA
Summary of Contents for WarmTouch
Page 1: ...Service Manual WarmTouch Convective Warming Unit...
Page 8: ...Page Left Intentionally Blank vi Service Manual...
Page 32: ...Page Left Intentionally Blank 1 18 Service Manual Introduction...
Page 42: ...Installation 3 4 Service Manual Figure 3 3 Power Cord Wrapped and Secured...
Page 45: ...Installing the Warming Unit Service Manual 3 7 Figure 3 5 Warming Unit Mounted on IV Pole...
Page 49: ...Installing the Warming Unit Service Manual 3 11 Figure 3 9 Warming Unit Mounted on Cart...
Page 50: ...Page Left Intentionally Blank 3 12 Service Manual Installation...
Page 70: ...Page Left Intentionally Blank 5 12 Service Manual Maintenance...
Page 84: ...Page Left Intentionally Blank 6 14 Service Manual Troubleshooting...
Page 105: ...Power Cord Replacement Service Manual 8 7 Figure 8 5 Power Cord Wrapped and Secured...
Page 118: ...Page Left Intentionally Blank 9 8 Service Manual Accessing the Inside of the Warming Unit...
Page 246: ...Enclosure Replacement 12 14 Service Manual Figure 12 10 Power Cord Wrapped and Secured...
Page 266: ...Page Left Intentionally Blank 12 34 Service Manual Enclosure Replacement...
Page 274: ...Page Left Intentionally Blank 13 8 Service Manual Product Specifications...
Page 313: ......