Installing the Warming Unit
Service Manual
3-5
3.4
Installing the Warming Unit
The warming unit can be used as a standalone device, or it can be mounted
on an IV pole, bed, or optional transport cart. The following sections provide
mounting instructions.
3.4.1
IV Pole Installation
The warming unit can be mounted on an IV pole with a maximum diameter of
3.2 cm (1.25 inches) and a minimum diameter of 1.9 cm (0.75 inches).
To avoid tipping, the warming unit must be mounted on the pole with the
handle no higher than 76 cm (30 inches) above the floor.
Requirements:
•
Measuring tape
To install the warming unit on an IV pole:
1.
On the IV pole, mark the maximum mounting height. The warming unit’s handle
must be no higher than 76 cm (30 inches) above the floor.
2.
Ensure that the power cord is secured at the back of the warming unit. Reference
3.
Locate the mounting channel with two clamps at the back of the warming unit.
Do not use the mounting channel with one clamp. Reference
4.
Loosen the knobs on the two clamps and rotate the clamp feet away from the
mounting channel.
Summary of Contents for WarmTouch
Page 1: ...Service Manual WarmTouch Convective Warming Unit...
Page 8: ...Page Left Intentionally Blank vi Service Manual...
Page 32: ...Page Left Intentionally Blank 1 18 Service Manual Introduction...
Page 42: ...Installation 3 4 Service Manual Figure 3 3 Power Cord Wrapped and Secured...
Page 45: ...Installing the Warming Unit Service Manual 3 7 Figure 3 5 Warming Unit Mounted on IV Pole...
Page 49: ...Installing the Warming Unit Service Manual 3 11 Figure 3 9 Warming Unit Mounted on Cart...
Page 50: ...Page Left Intentionally Blank 3 12 Service Manual Installation...
Page 70: ...Page Left Intentionally Blank 5 12 Service Manual Maintenance...
Page 84: ...Page Left Intentionally Blank 6 14 Service Manual Troubleshooting...
Page 105: ...Power Cord Replacement Service Manual 8 7 Figure 8 5 Power Cord Wrapped and Secured...
Page 118: ...Page Left Intentionally Blank 9 8 Service Manual Accessing the Inside of the Warming Unit...
Page 246: ...Enclosure Replacement 12 14 Service Manual Figure 12 10 Power Cord Wrapped and Secured...
Page 266: ...Page Left Intentionally Blank 12 34 Service Manual Enclosure Replacement...
Page 274: ...Page Left Intentionally Blank 13 8 Service Manual Product Specifications...
Page 313: ......