AC Power Inlet Replacement
Service Manual
10-53
3.
Disconnect the AC power inlet ground cable from the equipotential stud as
follows (
):
a.
Remove the top nut from the threaded end of the stud.
b.
Remove the lock washer and lift the power supply ground cable terminal off
the stud.
c.
Remove the next nut from the stud.
d.
Remove the next lock washer and lift the AC power inlet ground cable termi-
nal off the stud.
e.
Leave the equipotential stud on the enclosure.
Figure 10-39.
Equipotential Stud Connections
4.
Remove the two screws securing the AC power inlet assembly to the enclosure,
and remove the assembly.
5.
Position the new AC power inlet assembly as shown in
page 10-52. Make sure that the ground-pin side of the assembly is toward the
bottom of the enclosure.
6.
Attach the AC power inlet to the enclosure with the two screws. Tighten the
screws to 0.4 to 0.6 N-m (3.5 to 5.3 lb-in).
1
Power Supply Ground Cable
(Nut, Lock Washer,Terminal)
2
AC Power Inlet Ground Cable
(Nut, Lock Washer, Terminal)
Summary of Contents for WarmTouch
Page 1: ...Service Manual WarmTouch Convective Warming Unit...
Page 8: ...Page Left Intentionally Blank vi Service Manual...
Page 32: ...Page Left Intentionally Blank 1 18 Service Manual Introduction...
Page 42: ...Installation 3 4 Service Manual Figure 3 3 Power Cord Wrapped and Secured...
Page 45: ...Installing the Warming Unit Service Manual 3 7 Figure 3 5 Warming Unit Mounted on IV Pole...
Page 49: ...Installing the Warming Unit Service Manual 3 11 Figure 3 9 Warming Unit Mounted on Cart...
Page 50: ...Page Left Intentionally Blank 3 12 Service Manual Installation...
Page 70: ...Page Left Intentionally Blank 5 12 Service Manual Maintenance...
Page 84: ...Page Left Intentionally Blank 6 14 Service Manual Troubleshooting...
Page 105: ...Power Cord Replacement Service Manual 8 7 Figure 8 5 Power Cord Wrapped and Secured...
Page 118: ...Page Left Intentionally Blank 9 8 Service Manual Accessing the Inside of the Warming Unit...
Page 246: ...Enclosure Replacement 12 14 Service Manual Figure 12 10 Power Cord Wrapped and Secured...
Page 266: ...Page Left Intentionally Blank 12 34 Service Manual Enclosure Replacement...
Page 274: ...Page Left Intentionally Blank 13 8 Service Manual Product Specifications...
Page 313: ......