Equipotential Stud Replacement
Service Manual
10-57
4.
From the bottom of the enclosure, insert the threaded end of the stud into the
hole (
).
Figure 10-42.
Equipotential Stud and Washer
5.
Connect the cables to the equipotential stud as follows (
a.
Place the AC power inlet ground cable terminal on the threaded end of the stud.
b.
Place a lock washer on top of the terminal.
c.
Place a nut on top of the lock washer. Tighten the nut to 3.2 to 3.8 N-m (28.3
to 33.6 lb-in).
d.
Place the power supply ground cable terminal on top of the nut.
e.
Place a lock washer on top of the terminal.
f.
Place a nut on top of the lock washer. Tighten the nut to 3.2 to 3.8 N-m (28.3
to 33.6 lb-in).
6.
Follow the procedure
Rejoining the Front and Rear Enclosures
7.
Perform the following tests:
a.
Run an ohmmeter or digital multi-meter with resistance function between the
ground pin on the AC power inlet and the equipotential stud. The value must
be <1 Ohm.
b.
c.
Summary of Contents for WarmTouch
Page 1: ...Service Manual WarmTouch Convective Warming Unit...
Page 8: ...Page Left Intentionally Blank vi Service Manual...
Page 32: ...Page Left Intentionally Blank 1 18 Service Manual Introduction...
Page 42: ...Installation 3 4 Service Manual Figure 3 3 Power Cord Wrapped and Secured...
Page 45: ...Installing the Warming Unit Service Manual 3 7 Figure 3 5 Warming Unit Mounted on IV Pole...
Page 49: ...Installing the Warming Unit Service Manual 3 11 Figure 3 9 Warming Unit Mounted on Cart...
Page 50: ...Page Left Intentionally Blank 3 12 Service Manual Installation...
Page 70: ...Page Left Intentionally Blank 5 12 Service Manual Maintenance...
Page 84: ...Page Left Intentionally Blank 6 14 Service Manual Troubleshooting...
Page 105: ...Power Cord Replacement Service Manual 8 7 Figure 8 5 Power Cord Wrapped and Secured...
Page 118: ...Page Left Intentionally Blank 9 8 Service Manual Accessing the Inside of the Warming Unit...
Page 246: ...Enclosure Replacement 12 14 Service Manual Figure 12 10 Power Cord Wrapped and Secured...
Page 266: ...Page Left Intentionally Blank 12 34 Service Manual Enclosure Replacement...
Page 274: ...Page Left Intentionally Blank 13 8 Service Manual Product Specifications...
Page 313: ......