Power Supply Replacement
Service Manual
10-45
3.
Remove the four screws securing the power supply to the enclosure.
Caution:
The institution should follow local governing ordinances and recycling
instructions regarding disposal or recycling of filter and device
components or end of life of the product.
4.
Dispose of the power supply as required by WEEE Directive.
5.
Position the new power supply on the four mounting bosses, with the GND ter-
minal at the lower right near the AC power inlet (
6.
Attach the power supply to the mounting bosses with the four screws. Tighten
the screws to 0.4 to 0.6 N-m (3.5 to 5.3 lb-in).
7.
Connect the cables to the new power supply as follows (
a.
Connect the ground cable from the equipotential stud to the GND terminal.
Make sure the connector is fully seated on the terminal, and that the heat
shrink on the adjacent capacitor does not get caught under the connector.
b.
Connect the AC power inlet wire assembly to the 3-pin CON1 connector.
Make sure the connector latch engages.
c.
Connect the power supply output cable to the 4-pin CON2 connector. Make
sure the connector latch engages.
Note:
If a cable tie previously secured the AC power inlet wire assembly to the
power supply mounting boss (see
), install a new cable tie at the same
position. The cable tie should cause the wires connected to CON1 to form an arch
approximately 4 cm (1.5 in) above the connector. Trim off the end of the cable tie.
8.
Follow the procedure
Rejoining the Front and Rear Enclosures
9.
Perform the following tests:
a.
b.
Summary of Contents for WarmTouch
Page 1: ...Service Manual WarmTouch Convective Warming Unit...
Page 8: ...Page Left Intentionally Blank vi Service Manual...
Page 32: ...Page Left Intentionally Blank 1 18 Service Manual Introduction...
Page 42: ...Installation 3 4 Service Manual Figure 3 3 Power Cord Wrapped and Secured...
Page 45: ...Installing the Warming Unit Service Manual 3 7 Figure 3 5 Warming Unit Mounted on IV Pole...
Page 49: ...Installing the Warming Unit Service Manual 3 11 Figure 3 9 Warming Unit Mounted on Cart...
Page 50: ...Page Left Intentionally Blank 3 12 Service Manual Installation...
Page 70: ...Page Left Intentionally Blank 5 12 Service Manual Maintenance...
Page 84: ...Page Left Intentionally Blank 6 14 Service Manual Troubleshooting...
Page 105: ...Power Cord Replacement Service Manual 8 7 Figure 8 5 Power Cord Wrapped and Secured...
Page 118: ...Page Left Intentionally Blank 9 8 Service Manual Accessing the Inside of the Warming Unit...
Page 246: ...Enclosure Replacement 12 14 Service Manual Figure 12 10 Power Cord Wrapped and Secured...
Page 266: ...Page Left Intentionally Blank 12 34 Service Manual Enclosure Replacement...
Page 274: ...Page Left Intentionally Blank 13 8 Service Manual Product Specifications...
Page 313: ......