Operation
4-6
Service Manual
Figure 4-3.
Warming System Set to 40°C (Medium)
Note that when you change the temperature setting, an amber arrow appears
next to the current temperature while it adjusts. When the temperature is
within approximately 1.5°C of the new setting, the arrow disappears.
4.6.1
Using Boost Mode
Boost mode allows rapid warming of patients. Air is generated at 47°C
(116.6°F) for 45 minutes. After 45 minutes, the setting automatically changes
to 45°C (High).
To use Boost mode, press the
47°C (Boost Mode)
key. A countdown timer
shows the time remaining before the temperature switches to the 45°C (High)
setting. Boost mode sessions can be repeated as necessary.
Figure 4-4.
Warming System Set to Boost Mode
Summary of Contents for WarmTouch
Page 1: ...Service Manual WarmTouch Convective Warming Unit...
Page 8: ...Page Left Intentionally Blank vi Service Manual...
Page 32: ...Page Left Intentionally Blank 1 18 Service Manual Introduction...
Page 42: ...Installation 3 4 Service Manual Figure 3 3 Power Cord Wrapped and Secured...
Page 45: ...Installing the Warming Unit Service Manual 3 7 Figure 3 5 Warming Unit Mounted on IV Pole...
Page 49: ...Installing the Warming Unit Service Manual 3 11 Figure 3 9 Warming Unit Mounted on Cart...
Page 50: ...Page Left Intentionally Blank 3 12 Service Manual Installation...
Page 70: ...Page Left Intentionally Blank 5 12 Service Manual Maintenance...
Page 84: ...Page Left Intentionally Blank 6 14 Service Manual Troubleshooting...
Page 105: ...Power Cord Replacement Service Manual 8 7 Figure 8 5 Power Cord Wrapped and Secured...
Page 118: ...Page Left Intentionally Blank 9 8 Service Manual Accessing the Inside of the Warming Unit...
Page 246: ...Enclosure Replacement 12 14 Service Manual Figure 12 10 Power Cord Wrapped and Secured...
Page 266: ...Page Left Intentionally Blank 12 34 Service Manual Enclosure Replacement...
Page 274: ...Page Left Intentionally Blank 13 8 Service Manual Product Specifications...
Page 313: ......