Product Specifications
13-2
Service Manual
13.3
Electrical Requirements
13.4
Environmental Specifications
13.4.1
Operating
13.4.2
Shipping and Storage
Note:
Before operating the warming unit after it has been shipped or stored, allow the unit
to acclimate to the operating environment for at least 60 minutes.
Power Requirements
100 to 240 V AC
Max. current at 100 V = 8 A
Max. current at 240 V = 5 A
Input Frequency
50/60 Hz ±1 Hz
Fuse (x2)
Littlefuse (mfr.) 0218010; 250 V, 10 A
Temperature
18°C to 28°C (64.4°F to 82.4°F)
Relative Humidity
15% to 85% (non-condensing)
Temperature
-40ºC to 70ºC (-40ºF to 158ºF)
Relative Humidity
10% to 95% (non-condensing)
Atmospheric Pressure
12 kPa to 106 kPa (90 mmHg to 795 mmHg)
Summary of Contents for WarmTouch
Page 1: ...Service Manual WarmTouch Convective Warming Unit...
Page 8: ...Page Left Intentionally Blank vi Service Manual...
Page 32: ...Page Left Intentionally Blank 1 18 Service Manual Introduction...
Page 42: ...Installation 3 4 Service Manual Figure 3 3 Power Cord Wrapped and Secured...
Page 45: ...Installing the Warming Unit Service Manual 3 7 Figure 3 5 Warming Unit Mounted on IV Pole...
Page 49: ...Installing the Warming Unit Service Manual 3 11 Figure 3 9 Warming Unit Mounted on Cart...
Page 50: ...Page Left Intentionally Blank 3 12 Service Manual Installation...
Page 70: ...Page Left Intentionally Blank 5 12 Service Manual Maintenance...
Page 84: ...Page Left Intentionally Blank 6 14 Service Manual Troubleshooting...
Page 105: ...Power Cord Replacement Service Manual 8 7 Figure 8 5 Power Cord Wrapped and Secured...
Page 118: ...Page Left Intentionally Blank 9 8 Service Manual Accessing the Inside of the Warming Unit...
Page 246: ...Enclosure Replacement 12 14 Service Manual Figure 12 10 Power Cord Wrapped and Secured...
Page 266: ...Page Left Intentionally Blank 12 34 Service Manual Enclosure Replacement...
Page 274: ...Page Left Intentionally Blank 13 8 Service Manual Product Specifications...
Page 313: ......