Internal Component Replacement (Inside Front Enclosure)
10-20
Service Manual
To replace the ribbon cable:
1.
Follow the procedure
Separating the Front and Rear Enclosures
2.
On the UI PCBA, open the locking ears at the P204 connector, and disconnect the
ribbon cable (
Figure 10-16.
Ribbon Cable Connections
3.
On the Power PCBA, open the locking ears at the P101 connector, and disconnect
the ribbon cable.
4.
Connect the new ribbon cable to the P101 connector on the Power PCBA. Make
sure the locking ears close completely on the connector.
5.
Connect the ribbon cable to the P204 connector on the UI PCBA. Make sure the
locking ears close completely on the connector.
1
UI PCBA
3
Ribbon Cable
(P101 Connection)
2
Ribbon Cable
(P204 Connection)
4
Power PCBA
Summary of Contents for WarmTouch
Page 1: ...Service Manual WarmTouch Convective Warming Unit...
Page 8: ...Page Left Intentionally Blank vi Service Manual...
Page 32: ...Page Left Intentionally Blank 1 18 Service Manual Introduction...
Page 42: ...Installation 3 4 Service Manual Figure 3 3 Power Cord Wrapped and Secured...
Page 45: ...Installing the Warming Unit Service Manual 3 7 Figure 3 5 Warming Unit Mounted on IV Pole...
Page 49: ...Installing the Warming Unit Service Manual 3 11 Figure 3 9 Warming Unit Mounted on Cart...
Page 50: ...Page Left Intentionally Blank 3 12 Service Manual Installation...
Page 70: ...Page Left Intentionally Blank 5 12 Service Manual Maintenance...
Page 84: ...Page Left Intentionally Blank 6 14 Service Manual Troubleshooting...
Page 105: ...Power Cord Replacement Service Manual 8 7 Figure 8 5 Power Cord Wrapped and Secured...
Page 118: ...Page Left Intentionally Blank 9 8 Service Manual Accessing the Inside of the Warming Unit...
Page 246: ...Enclosure Replacement 12 14 Service Manual Figure 12 10 Power Cord Wrapped and Secured...
Page 266: ...Page Left Intentionally Blank 12 34 Service Manual Enclosure Replacement...
Page 274: ...Page Left Intentionally Blank 13 8 Service Manual Product Specifications...
Page 313: ......