Maintenance
5-8
Service Manual
10.
Set the unit upright, and plug the power cord into the power outlet.
11.
Reset the filter counters as follows:
a.
Press the
On/Standby
key. The Replace Filter screen appears (
).
Figure 5-5.
Replace Filter Screen
b.
Press the
Replace Filter
key. The Filter Information screen appears (
).
Figure 5-6.
Filter Information Screen - Filter Expired
c.
Press the
Filter Reset
key.
d.
Press the key code as described in
on page 7-2. The filter
counters are reset to 365 days and 2000 hours.
e.
Press the
On/Standby
key to return to Standby mode. Press the
On/Standby
key again and verify that the Main screen appears (
) rather than the
Replace Filter screen.
Summary of Contents for WarmTouch
Page 1: ...Service Manual WarmTouch Convective Warming Unit...
Page 8: ...Page Left Intentionally Blank vi Service Manual...
Page 32: ...Page Left Intentionally Blank 1 18 Service Manual Introduction...
Page 42: ...Installation 3 4 Service Manual Figure 3 3 Power Cord Wrapped and Secured...
Page 45: ...Installing the Warming Unit Service Manual 3 7 Figure 3 5 Warming Unit Mounted on IV Pole...
Page 49: ...Installing the Warming Unit Service Manual 3 11 Figure 3 9 Warming Unit Mounted on Cart...
Page 50: ...Page Left Intentionally Blank 3 12 Service Manual Installation...
Page 70: ...Page Left Intentionally Blank 5 12 Service Manual Maintenance...
Page 84: ...Page Left Intentionally Blank 6 14 Service Manual Troubleshooting...
Page 105: ...Power Cord Replacement Service Manual 8 7 Figure 8 5 Power Cord Wrapped and Secured...
Page 118: ...Page Left Intentionally Blank 9 8 Service Manual Accessing the Inside of the Warming Unit...
Page 246: ...Enclosure Replacement 12 14 Service Manual Figure 12 10 Power Cord Wrapped and Secured...
Page 266: ...Page Left Intentionally Blank 12 34 Service Manual Enclosure Replacement...
Page 274: ...Page Left Intentionally Blank 13 8 Service Manual Product Specifications...
Page 313: ......