Error Codes and Suggested Resolutions
Service Manual
B-11
158
Category:
UI
Alarm Priority:
Low
Description:
One of the following conditions occurred:
•
Temperature was more than ±1°C from the setpoint for at least 30 consecu-
tive seconds after the setpoint was reached.
•
Temperature did not reach the specified setpoint within 10 minutes.
Resolution:
Check for:
•
Blockages in the air intake, hose, and filter.
This error may also appear due to issues at the customer site:
•
Low mains voltage
•
Ambient temperature outside of specification.
159
Category:
UI
Alarm Priority:
Low
Description:
Temperature > 48°C for at least 30 consecutive seconds.
Resolution:
Replace the power PCBA. See
163
Category:
UI
Alarm Priority:
Low
Description:
Bad CRC on packet received from AD microprocessor.
Resolution:
Replace the ribbon cable. See
.
If this does not resolve the issue, try replacing the power PCBA. See
If the problem persists, try replacing the UI PCBA. See
This error may also appear due to electromagnetic interference from other devices
at the customer site. Advise the customer to try increasing the distance between
devices, repositioning the cabling, or plugging the devices into separate outlet
circuit branches. See
Electromagnetic Compatibility (EMC)
on page 13-4 for addi-
tional guidance.
Summary of Contents for WarmTouch
Page 1: ...Service Manual WarmTouch Convective Warming Unit...
Page 8: ...Page Left Intentionally Blank vi Service Manual...
Page 32: ...Page Left Intentionally Blank 1 18 Service Manual Introduction...
Page 42: ...Installation 3 4 Service Manual Figure 3 3 Power Cord Wrapped and Secured...
Page 45: ...Installing the Warming Unit Service Manual 3 7 Figure 3 5 Warming Unit Mounted on IV Pole...
Page 49: ...Installing the Warming Unit Service Manual 3 11 Figure 3 9 Warming Unit Mounted on Cart...
Page 50: ...Page Left Intentionally Blank 3 12 Service Manual Installation...
Page 70: ...Page Left Intentionally Blank 5 12 Service Manual Maintenance...
Page 84: ...Page Left Intentionally Blank 6 14 Service Manual Troubleshooting...
Page 105: ...Power Cord Replacement Service Manual 8 7 Figure 8 5 Power Cord Wrapped and Secured...
Page 118: ...Page Left Intentionally Blank 9 8 Service Manual Accessing the Inside of the Warming Unit...
Page 246: ...Enclosure Replacement 12 14 Service Manual Figure 12 10 Power Cord Wrapped and Secured...
Page 266: ...Page Left Intentionally Blank 12 34 Service Manual Enclosure Replacement...
Page 274: ...Page Left Intentionally Blank 13 8 Service Manual Product Specifications...
Page 313: ......