Error Codes
B-10
Service Manual
144
Category:
POST
Alarm Priority:
Low
Description:
UI calculated checkword does not match the stored checkword.
Resolution:
Replace the UI PCBA. See
145
Category:
POST
Alarm Priority:
Low
Description:
GPX calculated checkword does not match the stored checkword.
Resolution:
Replace the UI PCBA. See
152
Category:
UI
Alarm Priority:
Low
Description:
External watchdog timed out.
Resolution:
Replace the UI PCBA. See
157
Category:
UI
Alarm Priority:
Medium
Description:
Thermostat open.
Resolution:
Check for:
•
Thermistor dislodged from hose duct adapter - See
.
If the thermistor is in position, replace the power PCBA. See
Summary of Contents for WarmTouch
Page 1: ...Service Manual WarmTouch Convective Warming Unit...
Page 8: ...Page Left Intentionally Blank vi Service Manual...
Page 32: ...Page Left Intentionally Blank 1 18 Service Manual Introduction...
Page 42: ...Installation 3 4 Service Manual Figure 3 3 Power Cord Wrapped and Secured...
Page 45: ...Installing the Warming Unit Service Manual 3 7 Figure 3 5 Warming Unit Mounted on IV Pole...
Page 49: ...Installing the Warming Unit Service Manual 3 11 Figure 3 9 Warming Unit Mounted on Cart...
Page 50: ...Page Left Intentionally Blank 3 12 Service Manual Installation...
Page 70: ...Page Left Intentionally Blank 5 12 Service Manual Maintenance...
Page 84: ...Page Left Intentionally Blank 6 14 Service Manual Troubleshooting...
Page 105: ...Power Cord Replacement Service Manual 8 7 Figure 8 5 Power Cord Wrapped and Secured...
Page 118: ...Page Left Intentionally Blank 9 8 Service Manual Accessing the Inside of the Warming Unit...
Page 246: ...Enclosure Replacement 12 14 Service Manual Figure 12 10 Power Cord Wrapped and Secured...
Page 266: ...Page Left Intentionally Blank 12 34 Service Manual Enclosure Replacement...
Page 274: ...Page Left Intentionally Blank 13 8 Service Manual Product Specifications...
Page 313: ......