27
developed, predetermined force zones, the device will adjust its firing speed in each
zone to optimize staple formation.
If the stapler slows down during firing before the lower clamp cover reaches the distal
end, it is likely the firing force reached the force zone threshold and moved into the
next force zone level and the device firing speed was slowed to optimize stapling
performance.
Incomplete Firing
If the stapler stops firing before the lower clamp cover reaches the distal end, it is
likely the firing force measured reached the upper design limit of the stapling reload.
Carefully assess the tissue to determine whether to attempt to complete the firing.
PRECAUTION
When dividing major vascular structures, adhere to the basic surgical principles
of proximal and distal control.
If the stapler stops while firing:
1. Release the DOWN/FIRE button, and any other button or toggle that may be
pressed.
2. Inspect the stapling reload for an obstruction, excessively thick tissue, or for
completion of firing.
3. If continued firing is desired, attempt to complete the firing by pressing and holding
the DOWN/FIRE button until completion of firing.
4. If the stapler is unable to complete the firing, press UP on the toggle to RETRACT the
knife of the reload to the fully clamped position. Press and hold UP/OPEN again to fully
open the jaws of the stapling reload.
5. If unsuccessful in removing the stapling reload from the tissue, refer to the
Instructions for Use section to remove the reload. If that fails, follow the instructions
described in the Alternate Opening Procedure or in the Manual Retraction Tool
Procedure.
Alternative Opening Procedures
If the stapler failed to remove the stapling reload from the tissue, the following
procedures may be used to attempt to retract the knife and open the jaws of the
reload.
The alternative opening procedures can be performed using the existing power
stapler or with a new power stapler if a replacement power handle and sterile control
Instructions for Use
Summary of Contents for Signia Stapling System
Page 1: ...User Manual Signia Stapling System en cs sk sl et...
Page 2: ......
Page 4: ......
Page 6: ......
Page 20: ...14...
Page 40: ...34...
Page 41: ...U ivatelsk p ru ka Signia Svorkovac syst m cs...
Page 42: ......
Page 43: ...Obsah 1 P ehled syst mu 35 2 N vod k pou it 49 3 i t n a sterilizace 67...
Page 44: ......
Page 58: ...48 P ehled syst mu...
Page 78: ...68...
Page 79: ...N vod na pou itie Signia Svorkovac syst m sk...
Page 80: ......
Page 81: ...Obsah 1 Preh ad syst mu 69 2 N vod na pou itie 83 3 istenie a steriliz cia 103...
Page 82: ......
Page 96: ...82 Preh ad syst mu...
Page 116: ...102 N vod na pou itie...
Page 118: ...104...
Page 119: ...Priro nik za uporabo Signia Spenjalni sistem sl...
Page 120: ......
Page 121: ...Kazalo vsebine 1 Pregled sistema 105 2 Navodila za uporabo 119 3 i enje in sterilizacija 139...
Page 122: ......
Page 136: ...118 Pregled sistema...
Page 156: ...138 Navodila za uporabo...
Page 158: ...140...
Page 159: ...Kasutus juhend Signia Klammerdamis s steem et...
Page 160: ......
Page 161: ...Sisukord 1 S steemi levaade 141 2 Kasutusjuhised 155 3 Puhastamine ja steriliseerimine 175...
Page 162: ......
Page 176: ...154 S steemi levaade...
Page 196: ...174 Kasutusjuhised...
Page 198: ......