Using Oximeter Softkey Menus
Operator’s Manual
43
modified from the power-on default values. See
Adjusting the Factory Default Settings
a.
Upper and Lower SpO
2
Limits
—The neonate default upper limit is 95% and the
lower limit is 85%. An alarm sounds each time patient saturation violates these alarm
limits.
b.
SatSeconds Alarm Management
—The neonate default SatSeconds value is OFF.
Reference
Using the SatSeconds™ Alarm Management Feature
c.
OxiMax SPD™ Alert (SPD) Feature
—This option is not available for neonates.
Reference
Using the OxiMax SPD™ Alert Feature
d.
Upper and Lower Pulse Rate Limits
—The neonate default upper limit is 190 bpm
and the lower limit is 90 bpm. An alarm sounds each time patient pulse rate violates
these alarm limits.
e.
Pulse Rate Delay Alarm Management
—The neonate default Pulse Rate Delay is OFF,
but can be set to a five (5) or ten (10) second pulse rate alarm delay. Reference to
the Pulse Rate Delay Alarm Management Feature
3.
ADULT softkey
—Use this softkey to set upper and lower limits for adult and pediatric
patients. Scroll through each limit setting option using the SELECT softkey until you reach
the limit value you wish to change. Change the value by pressing the ADJUST UP or
ADJUST DOWN key below the ALARM SILENCE key on the oximeter panel until you reach
the desired value. This setting only holds until power recycle. Decimal points to the right
of the displayed values indicate those limits have been modified from the power-on
default values. See
Adjusting the Factory Default Settings
a.
Upper and Lower SpO
2
Limits
—The adult default upper limit is 100% and the lower
limit is 85%. An alarm sounds each time patient saturation violates these alarm limits.
b.
SatSeconds Alarm Management
—The adult default SatSeconds value is 100
SatSeconds. For more information, refer to
Using the SatSeconds™ Alarm Management
c.
OxiMax SPD™ Alert (SPD) Feature
—The SPD feature is not indicated for use on
pediatric patients, but on adult patients only. The adult default SPD alarm sensitivity is
a value of one (1) for most sensitive to patterns of desaturation. For more information,
refer to
Using the OxiMax SPD™ Alert Feature
, page 54. When SPD is enabled, the
SatSeconds feature is automatically enabled with a setting of 100.
d.
Upper and Lower Pulse Rate Limits
—The adult default upper limit is 170 bpm and
the lower limit is 40 bpm. An alarm sounds each time patient pulse rate violates these
alarm limits.
e.
Pulse Rate Delay Alarm Management
—The adult default Pulse Rate Delay is OFF,
but can be set to a five (5) or ten (10) second pulse rate alarm delay. For more
information, refer to
Using the Pulse Rate Delay Alarm Management Feature
TREND Menu
Choose the type of trend data to view by selecting trend data from the oximeter
(MONITR) or trend data from the pulse oximetry sensor (SENSOR) in the Trend menu.
Summary of Contents for Nellcor OxiMax N-600x
Page 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Page 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Page 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Page 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Page 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Page 145: ......