Reducing EMI (Electromagnetic Interference)
Operator’s Manual
25
1. Firmly connect a DOC-10 pulse oximetry cable to the oximeter’s SpO
2
Sensor Port.
Figure 13.
Front Panel Sensor Port Connector
2. Open the plastic latch at the other end of the DOC-10 pulse oximetry cable.
Figure 14.
Insertion of Cable into Plastic Latch on DOC-10 Pulse Oximetry Cable
3. Plug the cable and a Nellcor OxiMax SpO
2
sensor together.
4. Snap the plastic latch down over the connectors.
5. When the oximeter detects a valid pulse, it enters the monitoring mode and displays real-
time patient data.
6. Apply the sensor to the patient. Be sure to read the
Directions for Use
accompanying the
sensor.
Reducing EMI (Electromagnetic Interference)
This device has been tested and found to comply with the limits for medical devices related
to IEC 60601-1: 1988 + A1: 1991 + A2: 1995, EN60601-1:1990 + A11: 1993 + A12: 1993 + A13:
1996, and the Medical Device Directive 93/42/EEC. These limits are designed to provide
reasonable protection against harmful interference in a typical medical installation.
Because of the proliferation of radio frequency transmitting equipment and other
sources of electrical noise in health care environments (for example, electrosurgical
units, cellular phones, mobile two-way radios, electrical appliances, and
high-definition television), it is possible that high levels of such interference due to
close proximity or strength of a source may result in disruption of N-600x pulse
oximeter performance.
The OxiMax N-600x pulse oximeter is designed for use in environments in which the
pulse can be obscured by electromagnetic interference. During such interference,
measurements may seem inappropriate or the oximeter may not seem to operate
correctly. Disruption may be evidenced by erratic readings, cessation of operation, or
Summary of Contents for Nellcor OxiMax N-600x
Page 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Page 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Page 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Page 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Page 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Page 145: ......