Safety Tests
Operator’s Manual
129
Leakage Current
The following tables display the maximum earth and enclosure leakage current
allowed, as well as patient leakage.
Table 25.
Earth and Enclosure Leakage Current Specifications
Earth Leakage Current
Condition
AC Polarity
Line Cord
Neutral
Line Cord
IEC 60601-1
UL 60601-1
Normal
Normal
Closed
Closed
500
µA
300
µA
Single Fault
Open
Closed
1000
µA
Closed
Open
Normal
Reversed
Closed
Closed
500
µA
300
µA
Single Fault
Open
Closed
1000
µA
Closed
Open
Enclosure Leakage Current
Condition
AC Line Polarity
Neutral
Line Cord
Power Line
Ground
IEC 60601-1
UL 60601-1
Normal
Normal
Closed
Closed
100
µA
Single Fault
Open
Closed
500
µA
Closed
Open
Normal
Reversed
Closed
Closed
100
µA
Single Fault
Open
Closed
500
µA
Closed
Open
Summary of Contents for Nellcor OxiMax N-600x
Page 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Page 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Page 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Page 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Page 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Page 145: ......