Primary Speaker Failure
Operator’s Manual
107
Primary Speaker Failure
WARNING
If an OxiMax N-600x pulse oximeter reports a primary speaker failure, do not use the
oximeter longer than necessary to ensure patient safety. Contact a qualified service
technician, your local Nellcor representative, or Nellcor's Technical Services
Department at 1.800.635.5267 for assistance.
Table 17.
EEE Error Codes
Error
Code
Error Message
Corrective Action
80
DEFAULTS LOST
The current power-on default settings have been lost and returned to
factory defaults. Qualified service personnel can use the
N-600x Service
Manual
to restore the desired power-on default settings.
81
SETTINGS LOST
The current settings (for example, alarm limits, alarm and pulse beep
volumes, alarm silence duration) have been lost and returned to
power-on defaults. Turn the oximeter off and back on again. If it is
necessary to have settings different from the power-on default settings,
turn the oximeter off and back on again, and reenter the desired
settings.
82
CLOCK SETTING
LOST
Date and time settings are lost. Reenter the date and time.
Recharge or replace the battery.
515, 518,
534, 535,
569
N-600x Boot
Version x.x.x.x
The application software is missing or corrupt. Notify a qualified service
technician.
529, 729
LOW BATTERY
The battery is discharged to a critically low level. The oximeter will
shutdown after 10 seconds.
Verify the SUPPLY VOLTAGE SELECTOR switch on the rear panel is set to
the proper voltage.
Connect the oximeter to AC power and turn back on. A warning
message is displayed and a low priority audible alarm sounded. Press
the ALARM SILENCE key twice before the oximeter can be used for
patient monitoring.
575
TRENDS LOST
Oximeter trends are corrupt and will be cleared. Turn off the oximeter,
and then back on again.
701-716,
720-724,
732-740,
576-582
POWER SUPPLY
FAILURE
The oximeter power supply has detected an error. The oximeter will
shutdown after 10 seconds. Verify the oximeter is operating within
specified environmental conditions. Notify a qualified service
technician.
717, 718
BATTERY FAILURE
The oximeter has detected a battery open or short condition. The
oximeter will shutdown after 10 seconds. Battery should be replaced.
Notify a qualified service technician.
725-728,
730
REPLACE BATTERY
The battery is not charging properly. The oximeter will shutdown after
10 seconds. Battery should be replaced. Notify a qualified service
technician.
Summary of Contents for Nellcor OxiMax N-600x
Page 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Page 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Page 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Page 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Page 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Page 145: ......