Managing OxiMax™ Pulse Oximetry Sensor Data
Operator’s Manual
69
•
BANDAGE ASSEMBLY
•
NAIL POLISH
•
SENSOR TOO TIGHT?
•
REPOSITION SENSOR
•
ISOLATE INTERFERENCE SOURCE
•
CLEAN SENSOR SITE
To clear messages, review the listed corrective actions and take appropriate steps to
clear the condition, then press the EXIT softkey. Once closed, the sensor message
screen will not return until a new condition occurs.
Figure 58.
Corrective Action Screen
Enabling or Disabling the Sensor Message Feature
The OxiMax pulse oximetry sensor message setup display allows you to enable or
disable the sensor message feature. When disabled, neither the “SENSOR NOT
POSTING” nor the “SUGGESTED ACTION” messages display.
1. With the oximeter in the normal monitoring mode, press the SETUP softkey.
2. Press the SENSOR softkey.
3. Press the MSG softkey.
Figure 59.
Enable Sensor Message Screen
4. Press the ADJUST UP or ADJUST DOWN key below the ALARM SILENCE key on the
oximeter to toggle the ENABLE message.
5. Press the EXIT softkey.
Summary of Contents for Nellcor OxiMax N-600x
Page 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Page 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Page 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Page 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Page 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Page 145: ......