Operating the Oximeter
34
Operator’s Manual
•
SPD
feature
—Use the OxiMax SPD™ Alert (SPD) feature to detect patterns of desaturation in the
SpO
2
trend in adults. With the SPD feature enabled, the display view includes both the SatSeconds
circle icon and SPD triangle icon and their settings. The SPD alarm sensitivity value appears just
below the SPD icon. When the SPD feature is enabled, the triangle icon fills from the bottom to the
top as desaturation patterns develop. The triangle icon empties from the top to the bottom as
patterns dissipate. If the SPD icon reaches full, it indicates a threshold violation and the SPD alarm
sounds. The oximeter will sound the SPD alarm sooner if the SPD alarm sensitivity is set to the
default value of one (1). A less sensitive setting will result in less frequent alarms. Visible when the
SPD feature is activated and the SPD feature is enabled. For more information about activation,
contact Nellcor’s Technical Services Department at 1.800.635.5267 or your local Nellcor
representative. For more information about using the feature, reference
When the oximeter is not attached to AC power, it runs on an internal battery. When using the
internal battery, a battery fuel gauge appears in the upper left corner of the display.
To select the GCF Display
1. With the oximeter in the normal monitoring mode, press the SETUP softkey.
2. Press the VIEW softkey.
3. Press the NEXT softkey.
4. Press the GCF softkey.
Figure 19.
General Care Format Display View
Using the Pleth Display
Pleth Display Overview
Use this display view for visual monitoring information in waveform. The
plethysmographic (pleth) display includes a “wiper bar” plethysmographic
waveform, menu bar, and current measured SpO
2
and pulse rate, upper and lower
limit settings. It also includes a battery fuel gauge if running on battery power.
Plethysmographic waveforms with peak to peak amplitudes less than ten pulse
amplitude units (PAUs) are associated. Refer to
, page 115, for a
description of the pleth waveform. If enabled, SatSeconds and SPD icons display.
•
SatSeconds feature
—For mild or brief SpO
2
limit violations, use the SatSeconds feature
to reduce nuisance alarms. With the SatSeconds feature enabled, the SatSeconds circle
icon fills in the clockwise direction as the SatSeconds alarm management system detects
SpO
2
readings outside of the limit setting. The SatSeconds circle icon empties in
counterclockwise direction when SpO
2
readings are within limits. When the icon reaches
full, a medium priority alarm sounds. For more information about using the feature,
reference
Summary of Contents for Nellcor OxiMax N-600x
Page 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Page 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Page 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Page 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Page 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Page 145: ......